物理学 > 仪器与探测器
[提交于 2019年8月11日 (v1),最后修订于 2020年1月19日 (此版本, v9)]
标题:Measurement of the neutron beam profile of the Back-n white neutron facility at CSNS with a Micromegas detector
查看原文 PDF 中文 PDF (下批处理) EasyResearch 阅读摘要:The Back-n white neutron beam line, which uses back-streaming white neutrons from the spallation target of the China Spallation Neutron Source, is used for nuclear data measurements. A Micromegas-based neutron detector with two variants was specially developed to measure the beam spot distribution for this beam line. In this article, the design, fabrication, and characterization of the detector are described. The results of the detector performance tests are presented, which include the relative electron transparency, the gain and the gain uniformity, and the neutron beam profile reconstruction capability. The result of the first measurement of the Back-n neutron beam spot distribution is also presented.