标题:Harnessing bifurcations in tapping-mode atomic force microscopy to calibrate time-varying tip-sample force measurements
查看原文 PDF 中文 PDF (下批处理) EasyResearch 阅读摘要:Torsional harmonic cantilevers allow measurement of time varying tip-sample forces in tapping-mode atomic force microscopy. Accuracy of these force measurements is important for quantitative nanomechanical measurements. Here we demonstrate a method to convert the torsional deflection signals into a calibrated force waveform with the use of non-linear dynamical response of the tapping cantilever. Specifically the transitions between steady oscillation regimes are used to calibrate the torsional deflection signals.
提交历史
来自:Ozgur Sahin[v1] 2008年11月26日 星期三 00:00:00 UTC