标题:Position resolution and efficiency measurements with large scale Thin Gap Chambers for the super LHC
查看原文 PDF 中文 PDF (下批处理) EasyResearch 阅读摘要:New developments in Thin Gap Chambers (TGC) detectors to provide fast trigger and high precision muon tracking under sLHC conditions are presented. The modified detectors are shown to stand a high total irradiation dose equivalent to 6 Coulomb/cm of wire, without showing any deterioration in their performance. Two large (1.2 x 0.8 m^2) prototypes containing four gaps, each gap providing pad, strips and wires readout, with a total thickness of 50 mm, have been constructed. Their local spatial resolution has been measured in a 100 GeV/c muon test beam at CERN. At perpendicular incidence angle, single gap position resolution better than 60 microns has been obtained. For incidence angle of 20 degrees resolution of less than 100 micron was achieved. TGC prototypes were also tested under a flux of 10^5 Hz/cm^2 of 5.5-6.5 MeV neutrons, showing a high efficiency for cosmic muons detection.
提交历史
来自:Erez Etzion[v1] 2011年2月2日 星期三 00:00:00 UTC