物理学 > 等离子体物理
[提交于 2015年6月4日]
标题:Sensitivity of an image plate system in the XUV (60 eV < E < 900 eV)
查看原文 PDF 中文 PDF (下批处理) EasyResearch 阅读摘要:Phosphor imaging plates (IPs) have been calibrated and proven useful for quantitative x-ray imaging in the 1 to over 1000 keV energy range. In this paper we report on calibration measurements made at XUV energies in the 60 to 900 eV energy range using beamline 6.3.2 at the Advanced Light Source at Lawrence Berkeley National Laboratory. We measured a sensitivity of ~25 plus or minus 15 counts/pJ over the stated energy range which is compatible with the sensitivity of Si photodiodes that are used for time-resolved measurements. Our measurements at 900 eV are consistent with the measurements made by Meadowcroft et al. at ~1 keV.
提交历史
来自:Bruce Failor[v1] 2015年6月4日 星期四 00:00:00 UTC