标题:Optimal Surface Marker Locations for Tumor Motion Estimation in Lung Cancer Radiotherapy

作者:Bin Dong, Yan Jiang Graves, Xun Jia, Steve B. Jiang
查看论文《Optimal Surface Marker Locations for Tumor Motion Estimation in Lung Cancer Radiotherapy》的 PDF
查看原文 PDF 中文 PDF (下批处理) EasyResearch 阅读
摘要:Using fiducial markers on patient's body surface to predict the tumor location is a widely used approach in lung cancer radiotherapy. The purpose of this work is to propose an algorithm that automatically identifies a sparse set of locations on the patient's surface with the optimal prediction power for the tumor motion. The sparse selection of markers on the external surface and the assumed linear relationship between the marker motion and the internal tumor motion are represented by a prediction matrix. Such a matrix is determined by solving an optimization problem, where the objective function contains a sparsity term that penalizes the number of markers chosen on the patient's surface. The performance of our algorithm has been tested on realistic clinical data of four lung cancer patients. Thoracic 4DCT scans with 10 phases are used for the study. On a reference phase, a grid of points are casted on the patient's surface (except for patient's back) and propagated to other phases via deformable image registration of the corresponding CT images. Tumor locations at each phase are also manually delineated. We use 9 out of 10 phases of the 4DCT images to identify a small group of surface markers that are most correlated with the motion of the tumor, and find the prediction matrix at the same time. The 10th phase is then used to test the accuracy of the prediction. It is found that on average 6 to 7 surface markers are necessary to predict tumor locations with a 3D error of about 1mm. In addition, the selected marker locations lie closely in those areas where surface point motion has a high correlation with the tumor motion. Our method can automatically select sparse locations on patient's external surface and estimate a correlation matrix based on 4DCT, so that the selected surface locations can be used to place fiducial markers to optimally predict internal tumor motions.
主题: 医学物理 (physics.med-ph); 优化与控制 (math.OC)
引用方式: arXiv:1204.4719 [physics.med-ph]
  (或者 arXiv:1204.4719v1 [physics.med-ph] 对于此版本)
  https://doi.org/10.48550/arXiv.1204.4719
相关 DOI: https://doi.org/10.1088/0031-9155/57/24/8201

提交历史

来自:Bin Dong Dr.
[v1] 2012年4月19日 星期四 22:25:37 UTC
全文链接:

访问论文:

  • 查看原文 PDF
  • 中文 PDF (下批处理)
  • EasyResearch 阅读
  • TeX 源文件
  • arXiv 原文页面
查看许可

当前浏览上下文:

physics.med-ph
< 上一篇   |   下一篇 >
最新 | 近期 | 2012-04
切换浏览方式为:
physics
math.OC

参考文献与引用

  • Google Scholar
  • Semantic Scholar

BibTeX 格式的引用

EasyXiv 是独立运营的中文预印本站点,与 arXiv 及康奈尔大学无隶属关系。 论文元数据与原文来自 arXiv.org,感谢 arXiv 提供开放获取服务。
关于 · 帮助 · 投稿指南 · 版权 · arXiv 原站