物理学 > 仪器与探测器
[提交于 2012年4月28日]
标题:The lead-glass electromagnetic calorimeters for the magnetic spectrometers in Hall C at Jefferson Lab
查看原文 PDF 中文 PDF (下批处理) EasyResearch 阅读摘要:The electromagnetic calorimeters of the various magnetic spectrometers in Hall C at Jefferson Lab are presented. For the existing HMS and SOS spectrometers design considerations, relevant construction information, and comparisons of simulated and experimental results are included. The energy resolution of the HMS and SOS calorimeters is better than $\sigma/E \sim 6%/\sqrt E $, and pion/electron ($\pi/e$) separation of about 100:1 has been achieved in energy range 1 -- 5 GeV. Good agreement has been observed between the experimental and simulated energy resolutions, but simulations systematically exceed experimentally determined $\pi^-$ suppression factors by close to a factor of two. For the SHMS spectrometer presently under construction details on the design and accompanying GEANT4 simulation efforts are given. The anticipated performance of the new calorimeter is predicted over the full momentum range of the SHMS. Good electron/hadron separation is anticipated by combining the energy deposited in an initial (preshower) calorimeter layer with the total energy deposited in the calorimeter.
提交历史
来自:Hamlet Mkrtchyan Dr.[v1] 2012年4月28日 星期六 15:07:54 UTC