物理学 > 仪器与探测器
[提交于 2012年7月10日]
标题:Counting photons at low temperature with a streaming time-to-digital converter
查看原文 PDF 中文 PDF (下批处理) EasyResearch 阅读摘要:We present some aspects of photon counting to study scintillators at low temperatures. A time-to-digital converter (TDC) had been configured to acquire several-minute-long streams of data, simplifying the multiple photon counting coincidence technique. Results in terms of light yield and time structure of a ZnWO4 scintillator are comparable to those obtained with a fast digitizer. Streaming data also provides flexibility in analyzing the data, in terms of coincidence window between the channels, and acquisition window of individual channels. We discuss the effect of changing these parameters, and use them to confirm low-energy features in the spectra of the number of detected photons, such as the 60 keV line from 241Am in the ZnWO4 sample. We lastly use the TDC to study the transmission of the optical cryostat employed in these studies at various temperatures.
提交历史
来自:Philippe Di Stefano[v1] 2012年7月10日 星期二 19:35:19 UTC