天体物理学 > 天体物理仪器与方法
[提交于 2012年7月23日]
标题:Measurement of the Quantum Efficiency of Hamamatsu R8520 Photomultipliers at Liquid Xenon Temperature
查看原文 PDF 中文 PDF (下批处理) EasyResearch 阅读摘要:Vacuum ultraviolet light sensitive photomultiplier tubes directly coupled to liquid xenon are being used to efficiently detect the 178 nm scintillation light in a variety of liquid xenon based particle detectors. Good knowledge of the performance of these photomultipliers under cryogenic conditions is needed to properly characterize these detectors. Here, we report on measurements of the quantum efficiency of Hamamatsu R8520 photomultipliers, used in the XENON Dark Matter Experiments. The quantum efficiency measurements at room temperature agree with the values provided by Hamamatsu. At low temperatures, between 160K and 170K, the quantum efficiency increases by $\sim5-11$% relative to the room temperature values.
提交历史
来自:Bin Choi[v1] 2012年7月23日 星期一 16:03:41 UTC