定量生物学 > 定量方法
[提交于 2012年11月11日]
标题:Survey of Extra-Low Frequency and Very-Low Frequency Magnetic Fields in Cell Culture Incubators
查看原文 PDF 中文 PDF (下批处理) EasyResearch 阅读摘要:A typical cell culture CO2 incubator was probed in detail to document the pattern of 60-Hz magnetic fields (MFs) inside the unit, as well as the ability of the incubator to attenuate environmental MFs. Subsequently, a survey of 46 cell culture incubators was performed. The survey measured MFs outside and inside the incubators, the frequency spectrum between 5 and 2000 Hz, and variations over time of the 60-Hz MF. Our measurements show an uneven spatial distribution, reflecting electronic and electrical components hidden within the walls. Attenuation of environmental MFs varied between 18 % and 33 %, signalling easy penetration into the units. MF levels, frequency spectra and variations over time were very different from one unit to the next. All 46 incubators surveyed had an average field greater than 0.2 microT; among them, 39 (85 %) had an average field greater than 1 microT. There is substantial work to be done in improving control over the MF environment of in vitro experiments in bio-medicine, particularly if they involve cancer cells.
提交历史
来自:Paul Héroux[v1] 2012年11月11日 星期日 20:40:00 UTC