标题:A simulation study of a dual-plate in-room PET system for dose verification in carbon ion therapy

作者:Ze Chen, Zheng-Guo Hu, Guo-Qing Xiao, Jin-Da Chen, Xiu-Ling Zhang, Zhong-Yan Guo, Zhi-Yu Sun, Wen-Xue Huang, Jian-Song Wang
查看论文《A simulation study of a dual-plate in-room PET system for dose verification in carbon ion therapy》的 PDF
查看原文 PDF 中文 PDF (下批处理) EasyResearch 阅读
摘要:Carbon ion therapy have the ability to overcome the limitation of convertional radiotherapy due to its most energy deposition in selective depth, usually called Bragg peak, which results in increased biological effectiness. During carbon ion therapy, lots positron emitters such as $^{11}$C, $^{15}$O, $^{10}$C are generated in irradiated tissues by nuclear reactions. Immediately after patient irradiation, PET scanners can be used to measure the spatial distribution of positron emitters, which can track the carbon beam to the tissue. In this study, we designed and evaluated an dual-plate in-room PET scanner to monitor patient dose in carbon ion therapy, which is based on GATE simulation platform. A dual-plate PET is designed to avoid interference with the carbon beam line and with patient positioning. Its performance was compared with that of four-head and full-ring PET scanners. The dual-plate, four-head and full-ring PET scanners consisted of 30, 60, 60 detector modules, respectively, with a 36 cm distance between directly opposite detector modules for dose deposition measurements. Each detector module was consisted of a 24$\times$24 array of 2$\times2\times$18 mm$^{3}$ LYSO pixels coupled to a Hamamatsu H8500 PMT. To esitmate the production yield of positron emitters, a 10$\times15\times$15 cm$^{3}$ cuboid PMMA phantom was irradiated with 172, 200, 250 AMeV $^{12}$C beams. 3D images of the activity distribution of the three type scanners are produced by an iterative reconstruction algorithm. By comparing the longitudinal profile of positron emitters, measured along the carbon beam path, we concluded that the development of a dual-plate PET scanner is feasible to monitor the dose distribution for carbon ion therapy.
主题: 医学物理 (physics.med-ph); 生物物理 (physics.bio-ph)
引用方式: arXiv:1310.2469 [physics.med-ph]
  (或者 arXiv:1310.2469v1 [physics.med-ph] 对于此版本)
  https://doi.org/10.48550/arXiv.1310.2469
相关 DOI: https://doi.org/10.1088/1674-1137/38/8/088202

提交历史

来自:Ze Chen
[v1] 2015年6月17日 星期三 00:00:00 UTC
全文链接:

访问论文:

  • 查看原文 PDF
  • 中文 PDF (下批处理)
  • EasyResearch 阅读
  • TeX 源文件
  • arXiv 原文页面
查看许可

当前浏览上下文:

physics.med-ph
< 上一篇   |   下一篇 >
最新 | 近期 | 2015-06
切换浏览方式为:
physics
physics.bio-ph

参考文献与引用

  • Google Scholar
  • Semantic Scholar

BibTeX 格式的引用

EasyXiv 是独立运营的中文预印本站点,与 arXiv 及康奈尔大学无隶属关系。 论文元数据与原文来自 arXiv.org,感谢 arXiv 提供开放获取服务。
关于 · 帮助 · 投稿指南 · 版权 · arXiv 原站