物理学 > 仪器与探测器
[提交于 2015年12月9日]
标题:Large Magnetic Shielding Factor Measured by Nonlinear Magneto-optical Rotation
查看原文 PDF 中文 PDF (下批处理) EasyResearch 阅读摘要:A passive magnetic shield was designed and constructed for magnetometer tests for the future neutron electric dipole moment experiment at TRIUMF. The axial shielding factor of the magnetic shield was measured using a magnetometer based on non-linear magneto-optical rotation of the plane of polarized laser light upon passage through a paraffin-coated vapour cell containing natural Rb at room temperature. The laser was tuned to the Rb D1 line, near the $^{85}$Rb $F=2\rightarrow 2,3$ transition. The shielding factor was measured by applying an axial field externally and measuring the magnetic field internally using the magnetometer. The axial shielding factor was determined to be $(1.3\pm 0.1)\times 10^{7}$, from an applied axial field of 1.45~$\mu$T in the background of Earth's magnetic field.
提交历史
来自:Jeffery Martin[v1] 2015年12月9日 星期三 00:00:00 UTC