计算机科学 > 新兴技术
[提交于 2015年7月13日]
标题:A Batteryless Sensor ASIC for Implantable Bio-impedance Applications
查看原文 PDF 中文 PDF (下批处理) EasyResearch 阅读摘要:The measurement of the biological tissue's electrical impedance is an active research field that has attracted a lot of attention during the last decades. Bio-impedances are closely related to a large variety of physiological conditions; therefore, they are useful for diagnosis and monitoring in many medical applications. Measuring living tissues, however, is a challenging task that poses countless technical and practical problems, in particular if the tissues need to be measured under the skin. This paper presents a bio-impedance sensor ASIC targeting a battery-free, miniature size, implantable device, which performs accurate 4-point complex impedance extraction in the frequency range from 2 kHz to 2 MHz. The ASIC is fabricated in 150 nm CMOS, has a size of 1.22 mm x 1.22 mm and consumes 165 uA from a 1.8 V power supply. The ASIC is embedded in a prototype which communicates with, and is powered by an external reader device through inductive coupling. The prototype is validated by measuring the impedances of different combinations of discrete components, measuring the electrochemical impedance of physiological solution, and performing ex vivo measurements on animal organs. The proposed ASIC is able to extract complex impedances with around 1 Ohm resolution; therefore enabling accurate wireless tissue measurements.
提交历史
来自:Saul Rodriguez[v1] 2015年7月13日 星期一 10:35:42 UTC