定量生物学 > 定量方法
[提交于 2016年11月21日]
标题:CMOS based image cytometry for detection of phytoplankton in ballast water
查看原文 PDF 中文 PDF (下批处理) EasyResearch 阅读摘要:We introduce an image cytometer (I-CYT) for the analysis of phytoplankton in fresh and marine water environments. A linear quantification of cell numbers was observed covering several orders of magnitude using cultures of Tetraselmis and Nannochloropsis measured by autofluorescence in a laboratory environment. We assessed the functionality of the system outside the laboratory by phytoplankton quantification of samples taken from a marine water environment (Dutch Wadden Sea, The Netherlands) and a fresh water environment (Lake Ijssel, The Netherlands). The I-CYT was also employed to study the effects of two ballast water treatment systems (BWTS), based on chlorine electrolysis and UV sterilization, with the analysis including the vitality of the phytoplankton. For comparative study and benchmarking of the I-CYT, a standard flow cytometer was used. Our results prove a limit of detection (LOD) of 10 cells/ml with an accuracy between 0.7 and 0.5 log, and a correlation of 88.29% in quantification and 96.21% in vitality, with respect to the flow cytometry results.
提交历史
来自:Juan Miguel Pérez Rosas[v1] 2016年11月21日 星期一 16:35:41 UTC