标题:Reproducibility of Retinal Thickness Measurements across Spectral-Domain Optical Coherence Tomography Devices using Iowa Reference Algorithm

作者:Adnan Rashid, Sebastian M. Waldstein, Bianca S. Gerendas, Hrvoje Bogunovic, Andreas Wahle, Kyungmoo Lee, Kai Wang, Christian Simader, Michael D. Abramoff, Ursula Schmidt-Erfurth, Milan Sonka
查看论文《Reproducibility of Retinal Thickness Measurements across Spectral-Domain Optical Coherence Tomography Devices using Iowa Reference Algorithm》的 PDF
查看原文 PDF 中文 PDF (下批处理) EasyResearch 阅读
摘要:PURPOSE: Establishing and obtaining consistent quantitative indices of retinal thickness from a variety of clinically used Spectral-Domain Optical Coherence Tomography scanners. DESIGN: Retinal images from five Spectral-Domain Optical Coherence Tomography scanners were used to determine total retinal thickness with scanner-specific correction factors establishing consistency of thickness measurement across devices. PARTICIPANTS: 55 Fovea-centered Spectral-Domain Optical Coherence Tomography volumes from eleven subjects were analyzed, obtained from Cirrus HD-OCT, RS-3000, Heidelberg Spectralis, RTVue and Topcon2000, seven subjects with retinal diseases and four normal controls. METHOD: The Iowa Reference Algorithm measured total retinal thickness. Nonlinear model of total retinal thickness measurement comparisons was derived and used for device-specific comparisons. Bland-Altman plots and pairwise predictive equations yielded pairwise scanner-specific differences. Mendel test determined whether measurement biases were constant for each scanner pair. RESULTS: Mendel test revealed that all pairwise scanner differences of total retinal thickness were constant across the cohort (p=0.992). Individual measurements can thus be bias-corrected and the Iowa Reference Algorithm serve as a scanner-agnostic independent standard of total retinal thickness across the five tested SD-OCT scanners. CONCLUSIONS: Combination of the Iowa Reference Algorithm with scanner-specific bias correction yields cross-scanner consistency of total retinal thickness measurements, facilitating scanning-device independent quantitative assessment of total retinal thickness, longitudinal follow-up quantification without requiring patients to be imaged on the same scanner model, and allowing for multi-center studies with heterogeneous device utilization when using the Iowa Reference Algorithm.
评论: 14 pages, 5 tables, 5 figures
主题: 医学物理 (physics.med-ph)
引用方式: arXiv:1612.06442 [physics.med-ph]
  (或者 arXiv:1612.06442v1 [physics.med-ph] 对于此版本)
  https://doi.org/10.48550/arXiv.1612.06442

提交历史

来自:Milan Sonka
[v1] 2016年12月19日 星期一 22:16:56 UTC
全文链接:

访问论文:

  • 查看原文 PDF
  • 中文 PDF (下批处理)
  • EasyResearch 阅读
  • TeX 源文件
  • arXiv 原文页面
查看许可

当前浏览上下文:

physics.med-ph
< 上一篇   |   下一篇 >
最新 | 近期 | 2016-12
切换浏览方式为:
physics

参考文献与引用

  • Google Scholar
  • Semantic Scholar

BibTeX 格式的引用

EasyXiv 是独立运营的中文预印本站点,与 arXiv 及康奈尔大学无隶属关系。 论文元数据与原文来自 arXiv.org,感谢 arXiv 提供开放获取服务。
关于 · 帮助 · 投稿指南 · 版权 · arXiv 原站