摘要:Purpose: Implanted fiducial markers are often used in radiotherapy to facilitate accurate visualization and localization of tumors. Typically, such markers are used to aid daily patient positioning and to verify the target's position during treatment. This work introduces a novel, automated method for identifying fiducial markers in planar x-ray imaging. Methods: In brief, the method consists of automated filtration and reconstruction steps that generate 3D templates of marker positions. The normalized cross-correlation was the used to identify fiducial markers in projection images. To quantify the accuracy of the technique, a phantom study was performed. 75 pre-treatment CBCT scans of 15 pancreatic cancer patients were analyzed to test the automated technique under real life conditions, including several challenging scenarios for tracking fiducial markers. Results: In phantom and patient studies, the method automatically tracked visible marker clusters in 100% of projection images. For scans in which a phantom exhibited 0D, 1D, and 3D motion, the automated technique showed median errors of 39 $\mu$m, 53 $\mu$m, and 93 $\mu$m, respectively. Human precision was worse in comparison. Automated tracking was performed accurately despite the presence of other metallic objects. Additionally, transient differences in the cross-correlation score identified instances where markers disappeared from view. Conclusions: A novel, automated method for producing dynamic templates of fiducial marker clusters has been developed. Production of these templates automatically provides measurements of tumor motion that occurred during the CBCT scan that was used to produce them. Additionally, using these templates with intra-fractional images could potentially allow for more robust real-time target tracking in radiotherapy.
评论: 12 pages, 7 figures
主题: 医学物理 (physics.med-ph)
引用方式: arXiv:1703.03870 [physics.med-ph]
  (或者 arXiv:1703.03870v1 [physics.med-ph] 对于此版本)
  https://doi.org/10.48550/arXiv.1703.03870
期刊参考: Medical Physics 44(2), 2017
相关 DOI: https://doi.org/10.1002/mp.12073

提交历史

来自:Bernard Jones
[v1] 2017年3月10日 星期五 23:29:21 UTC
全文链接:

访问论文:

  • 查看原文 PDF
  • 中文 PDF (下批处理)
  • EasyResearch 阅读
  • TeX 源文件
  • arXiv 原文页面
查看许可

当前浏览上下文:

physics.med-ph
< 上一篇   |   下一篇 >
最新 | 近期 | 2017-03
切换浏览方式为:
physics

参考文献与引用

  • Google Scholar
  • Semantic Scholar

BibTeX 格式的引用

EasyXiv 是独立运营的中文预印本站点,与 arXiv 及康奈尔大学无隶属关系。 论文元数据与原文来自 arXiv.org,感谢 arXiv 提供开放获取服务。
关于 · 帮助 · 投稿指南 · 版权 · arXiv 原站