跳到正文
EasyXiv 检索
投稿 关于 登录

按回车检索 · 直接输入编号(如 1706.03762)可跳转到论文

物理学 > 仪器与探测器

arXiv:1707.03392 (physics)
[提交于 2017年7月10日]

标题:On the use of NaI scintillation for high stability nuclear decay rate measurements

作者:Scott D. Bergeson, Michael J. Ware, Jeremy Hawk
查看论文《On the use of NaI scintillation for high stability nuclear decay rate measurements》的 PDF
查看原文 PDF 中文 PDF (下批处理) EasyResearch 阅读
摘要:We demonstrate the linearity and stability of a gamma-ray scintillation detector comprised of a NaI(Tl) crystal and a scientific-grade CMOS camera. After calibration, this detector exhibits excellent linearity more than three decades of activity levels. Because the detector is not counting pulses, no dead-time correction is required. When high activity sources are brought into close proximity to the NaI crystal, several minutes are required for the scintillation to achieve a steady state. On longer time scales, we measure drifts of a few percent over several days. These instabilities have important implications for precision determinations of nuclear decay rate stability.
评论: Submitted to Nucl Instrum Methods Phys Res A
主题: 仪器与探测器 (physics.ins-det); 核实验 (nucl-ex)
引用方式: arXiv:1707.03392 [physics.ins-det]
  (或者 arXiv:1707.03392v1 [physics.ins-det] 对于此版本)
  https://doi.org/10.48550/arXiv.1707.03392

提交历史

来自:Scott Bergeson
[v1] 2017年7月10日 星期一 22:23:13 UTC
全文链接:

访问论文:

  • 查看原文 PDF
  • 中文 PDF (下批处理)
  • EasyResearch 阅读
  • TeX 源文件
  • arXiv 原文页面
查看许可

当前浏览上下文:

physics.ins-det
< 上一篇   |   下一篇 >
最新 | 近期 | 2017-07
切换浏览方式为:
physics
nucl-ex

参考文献与引用

  • Google Scholar
  • Semantic Scholar

BibTeX 格式的引用

EasyXiv 是独立运营的中文预印本站点,与 arXiv 及康奈尔大学无隶属关系。 论文元数据与原文来自 arXiv.org,感谢 arXiv 提供开放获取服务。
关于 · 帮助 · 投稿指南 · 版权 · arXiv 原站