定量生物学 > 定量方法
[提交于 2018年4月11日]
标题:Blood Glucose Measurement Based on Infra-Red Spectroscopy
查看原文 PDF 中文 PDF (下批处理) EasyResearch 阅读摘要:An algorithm based on PLS regression has been developed and optimized for measuring blood glucose level using the infra-red transmission spectrum of blood samples. A set of blood samples were tagged with their glucose concentration using an accurate invasive glucometer and analyzed using a standard FTIR spectrometer. Using the developed algorithm, the results of the FTIR spectroscopy of the samples were analyzed to find the glucose concentration in the samples. The obtained glucose concentration by the algorithm were in good agreement with the results obtained by the standard glucometer, and the mean estimation error was 7 mg/dL. This error is in the range of available commercial invasive meters.
提交历史
来自:Mehdi Fardmanesh Prof.[v1] 2018年4月11日 星期三 00:00:00 UTC