标题:Dose Delivery Verification

作者:S. Safai
查看论文《Dose Delivery Verification》的 PDF
查看原文 PDF 中文 PDF (下批处理) EasyResearch 阅读
摘要:This paper focuses on some dosimetry aspects of proton therapy and pencil beam scanning based on the experience accumulated at Paul Scherrer Institute(PSI). The basic formalism for absolute dosimetry in proton therapy is outlined and the two main techniques and equipment to perform the primary beam monitor chamber calibration are presented. Depth-dose curve and lateral beam width measurements are exposed and discussed in detail, with particular attention to the size of the ionization chamber and the characteristic of scintillating-CCD dosimetry systems, respectively. It is also explained how the angular-spatial distribution of individual pencil beams can be determined in practice. The equipment and the techniques for performing regularmachine-specific quality checks are focused on (i)output constancy checks, (ii)pencil beam position and size checks and (iii)beam energy checks. Finally, patient-specific verification is addressed.
评论: 10 pages, Presented at the CAS- CERN Accelerator School on Accelerators for Medical Application, V\"osendorf, Austria, 26 May - 5 June, 2015
主题: 医学物理 (physics.med-ph)
引用方式: arXiv:1804.08509 [physics.med-ph]
  (或者 arXiv:1804.08509v1 [physics.med-ph] 对于此版本)
  https://doi.org/10.48550/arXiv.1804.08509
期刊参考: CERN Yellow Report CERN-2017-004-SP pp. 49-58
相关 DOI: https://doi.org/10.23730/CYRSP-2017-001.49

提交历史

来自:Scientific Information Service CERN
[v1] 2018年4月23日 星期一 15:31:28 UTC
全文链接:

访问论文:

  • 查看原文 PDF
  • 中文 PDF (下批处理)
  • EasyResearch 阅读
  • TeX 源文件
  • arXiv 原文页面
查看许可

当前浏览上下文:

physics.med-ph
< 上一篇   |   下一篇 >
最新 | 近期 | 2018-04
切换浏览方式为:
physics

参考文献与引用

  • Google Scholar
  • Semantic Scholar

BibTeX 格式的引用

EasyXiv 是独立运营的中文预印本站点,与 arXiv 及康奈尔大学无隶属关系。 论文元数据与原文来自 arXiv.org,感谢 arXiv 提供开放获取服务。
关于 · 帮助 · 投稿指南 · 版权 · arXiv 原站