跳到正文
EasyXiv 检索
投稿 关于 登录

按回车检索 · 直接输入编号(如 1706.03762)可跳转到论文

天体物理学 > 天体物理仪器与方法

arXiv:1810.00598 (astro-ph)
[提交于 2019年4月4日]

标题:Kinetic Inductance Detectors for the OLIMPO experiment: design and pre-flight characterization

作者:A. Paiella, A. Coppolecchia, L. Lamagna, P. A. R. Ade, E. S. Battistelli, M. G. Castellano, I. Colantoni, F. Columbro, G. D'Alessandro, P. de Bernardis, S. Gordon, S. Masi, P. Mauskopf, G. Pettinari, F. Piacentini, G. Pisano, G. Presta, C. Tucker
查看论文《Kinetic Inductance Detectors for the OLIMPO experiment: design and pre-flight characterization》的 PDF
查看原文 PDF 中文 PDF (下批处理) EasyResearch 阅读
摘要:We designed, fabricated, and characterized four arrays of horn--coupled, lumped element kinetic inductance detectors (LEKIDs), optimized to work in the spectral bands of the balloon-borne OLIMPO experiment. OLIMPO is a 2.6 m aperture telescope, aimed at spectroscopic measurements of the Sunyaev-Zel'dovich (SZ) effect. OLIMPO will also validate the LEKID technology in a representative space environment. The corrected focal plane is filled with diffraction limited horn-coupled KID arrays, with 19, 37, 23, 41 active pixels respectively at 150, 250, 350, and 460$\:$GHz. Here we report on the full electrical and optical characterization performed on these detector arrays before the flight. In a dark laboratory cryostat, we measured the resonator electrical parameters, such as the quality factors and the electrical responsivities, at a base temperature of 300$\:$mK. The measured average resonator $Q$s are 1.7$\times{10^4}$, 7.0$\times{10^4}$, 1.0$\times{10^4}$, and 1.0$\times{10^4}$ for the 150, 250, 350, and 460$\:$GHz arrays, respectively. The average electrical phase responsivities on resonance are 1.4$\:$rad/pW, 1.5$\:$rad/pW, 2.1$\:$rad/pW, and 2.1$\:$rad/pW; the electrical noise equivalent powers are 45$\:\rm{aW/\sqrt{Hz}}$, 160$\:\rm{aW/\sqrt{Hz}}$, 80$\:\rm{aW/\sqrt{Hz}}$, and 140$\:\rm{aW/\sqrt{Hz}}$, at 12 Hz. In the OLIMPO cryostat, we measured the optical properties, such as the noise equivalent temperatures (NET) and the spectral responses. The measured NET$_{\rm RJ}$s are $200\:\mu\rm{K\sqrt{s}}$, $240\:\mu\rm{K\sqrt{s}}$, $240\:\mu\rm{K\sqrt{s}}$, and $\:340\mu\rm{K\sqrt{s}}$, at 12 Hz; under 78, 88, 92, and 90 mK Rayleigh-Jeans blackbody load changes respectively for the 150, 250, 350, and 460 GHz arrays. The spectral responses were characterized with the OLIMPO differential Fourier transform spectrometer (DFTS) up to THz frequencies, with a resolution of 1.8 GHz.
评论: Published on JCAP
主题: 天体物理仪器与方法 (astro-ph.IM); 仪器与探测器 (physics.ins-det)
引用方式: arXiv:1810.00598 [astro-ph.IM]
  (或者 arXiv:1810.00598v1 [astro-ph.IM] 对于此版本)
  https://doi.org/10.48550/arXiv.1810.00598
相关 DOI: https://doi.org/10.1088/1475-7516/2019/01/039

提交历史

来自:Alessandro Paiella
[v1] 2019年4月4日 星期四 00:00:00 UTC
全文链接:

访问论文:

  • 查看原文 PDF
  • 中文 PDF (下批处理)
  • EasyResearch 阅读
  • TeX 源文件
  • arXiv 原文页面
查看许可

当前浏览上下文:

astro-ph.IM
< 上一篇   |   下一篇 >
最新 | 近期 | 2019-04
切换浏览方式为:
astro-ph
physics.ins-det

参考文献与引用

  • Google Scholar
  • Semantic Scholar

BibTeX 格式的引用

EasyXiv 是独立运营的中文预印本站点,与 arXiv 及康奈尔大学无隶属关系。 论文元数据与原文来自 arXiv.org,感谢 arXiv 提供开放获取服务。
关于 · 帮助 · 投稿指南 · 版权 · arXiv 原站