标题:Multiplexing Capabilities of Cold-Neutron Triple-Axis Spectrometer SIKA
查看原文 PDF 中文 PDF (下批处理) EasyResearch 阅读摘要:SIKA, a high-flux cold-neutron triple-axis spectrometer at the OPAL reactor at the Australian Nuclear Science and Technology Organization, is equipped with a 13-blade analyser and position-sensitive detector. This multiplexing design endows SIKA high flexibility to run in both traditional triple-axis and multiplexing analyser modes. In this study, two different multiplexing modes on SIKA are simulated using Monte-Carlo ray-tracing methods. The simulation results demonstrate SIKA capabilities to work in these operational modes, especially, the multi-Q const-Ef mode. This capability was demonstrated by measuring the phonon dispersion of a Pb single-crystal sample with the multi-Q const-Ef mode on SIKA. Compared to the traditional and multi-analyser triple-axis spectrometers, multiplexing modes on SIKA combine the advantages of the high data-acquisition efficiency and flexibility to focus on local areas of interest in the (Q, w) space.
提交历史
来自:Guochu Deng[v1] 2020年2月20日 星期四 00:00:00 UTC