电气工程与系统科学 > 信号处理
[提交于 2024年10月30日]
标题:Magneto-Inductive Powering and Uplink of In-Body Microsensors: Feasibility and High-Density Effects
查看原文 PDF 中文 PDF (下批处理) EasyResearch 阅读摘要:This paper studies magnetic induction for wireless powering and the data uplink of microsensors, in particular for future medical in-body applications. We consider an external massive coil array as power source (1 W) and data sink. For sensor devices at 12 cm distance from the array, e.g. beneath the human skin, we compute a minimum coil size of 150 um assuming 50 nW required chip activation power and operation at 750 MHz. A 275 um coil at the sensor allows for 1 Mbit/s uplink rate. Moreover, we study resonant sensor nodes in dense swarms, a key aspect of envisioned biomedical applications. In particular, we investigate the occurring passive relaying effect and cooperative transmit beamforming in the uplink. We show that the frequency- and location-dependent signal fluctuations in such swarms allow for significant performance gains when utilized with adaptive matching, spectrally-aware signaling and node cooperation. The work is based on a general magneto-inductive MIMO system model, which is introduced first.
提交历史
来自:Gregor Dumphart[v1] 2024年10月30日 星期三 00:00:00 UTC