物理学 > 仪器与探测器
[提交于 2019年5月3日]
标题:Linearized and Compensated Interferometric System for High-Velocity Traceable Length Calibration on a Metre Scale
查看原文 PDF 中文 PDF (下批处理) EasyResearch 阅读摘要:We report on a traceable calibration system for a 3500mm-long console that carries a measurement system for inspecting the diameter of a circular reactor chassis. The system uses two single-pass laser interferometers with homodyne fringe detection for measurement in two degrees of freedom. The hybrid FPGA-microcontroller control module carries out the fringe detection together with the application-specific scale linearization approach and the compensation of environmental influences such as thermal elongation and the refractive index of air fluctuations. We demonstrated the system feasibility with an accuracy of a few microns and translation velocity higher than 0:1 metre per second.
提交历史
来自:Simon Rerucha[v1] 2019年5月3日 星期五 00:00:00 UTC