标题:Active Personal Eye Lens Dosimetry with the Hybrid Pixelated Dosepix Detector
查看中文 PDF 查看原文 PDF EasyResearch 阅读摘要:Eye lens dosimetry has been an important field of research in the last decade. Dose measurements with a prototype of an active personal eye lens dosemeter based on the Dosepix detector are presented. The personal dose equivalent at $3\,$mm depth of soft tissue, $H_\text{p}(3)$, was measured in the center front of a water-filled cylinder phantom with a height and diameter of $20\,$cm. The energy dependence of the normalized response is investigated for mean photon energies between $12.4\,$keV and $248\,$keV for continuous reference radiation fields (N-series) according to ISO 4037. The response normalized to N-60 ($\overline{E}=47.9\,\text{keV}$) at $0^\circ$ angle of irradiation stays within the approval limits of IEC 61526 for angles of incidence between $-75^\circ$ and $+75^\circ$. Performance in pulsed photon fields was tested for varying dose rates from $0.1\,\frac{\text{Sv}}{\text{h}}$ up to $1000\,\frac{\text{Sv}}{\text{h}}$ and pulse durations from $1\,\text{ms}$ up to $10\,\text{s}$. The dose measurement works well within the approval limits (acc. to IEC 61526) up to $1\,\frac{\text{Sv}}{\text{h}}$. No significant influence of the pulse duration on the measured dose is found. Reproducibility measurements yield a coefficient of variation which does not exceed $1\,\%$ for two tested eye lens dosemeter prototypes.