电气工程与系统科学 > 信号处理
[提交于 2025年1月31日]
标题:Channel Characterization of Implantable Intrabody Communication through Experimental Measurements
查看原文 PDF 中文 PDF (下批处理) EasyResearch 阅读摘要:Intrabody communication (IBC), is a promising technology that can be utilized for data transmission across the human body. In this study, a galvanic coupled (GC)-based IBC channel has been investigated for implantable configuration both theoretically and experimentally in the frequency range of 0 to 2.5 MHz. Theoretical studies were performed by using finite element method (FEM) based simulation software, called Comsol Multiphysics. A cylindrical human arm was modeled with realistic values. Experimental studies were carried out with chicken breast tissue as a substitute for human tissue. The pseudorandom noise (PN) sequences were transmitted to investigate the correlative channel sounder of tissue model. Results showed that the frequency affects signal propagation through the tissue model. Additionally, it is crucial to cancel common-mode noise in the IBC channel to enhance communication quality.
提交历史
来自:Pietro Savazzi[v1] 2025年1月31日 星期五 00:00:00 UTC