标题:Laboratory evaluation of a wearable instrumented headband for rotational head kinematics measurement

作者:Anu Tripathi, Yang Wan, Sushant Malave, Sheila Turcsanyi, Alice Lux Fawzi, Alison Brooks, Haneesh Kesari, Traci Snedden, Peter Ferrazzano, Christian Franck, Rika Carlsen
查看论文《Laboratory evaluation of a wearable instrumented headband for rotational head kinematics measurement》的 PDF
查看原文 PDF 中文 PDF (下批处理) EasyResearch 阅读
摘要:Mild traumatic brain injuries (mTBI) are a highly prevalent condition with heterogeneous outcomes between individuals. A key factor governing brain tissue deformation and the risk of mTBI is the rotational kinematics of the head. Instrumented mouthguards are a widely accepted method for measuring rotational head motions, owing to their robust sensor-skull coupling. However, wearing mouthguards is not feasible in all situations, especially for long-term data collection. Therefore, alternative wearable devices are needed. In this study, we present an improved design and data processing scheme for an instrumented headband. Our instrumented headband utilizes an array of inertial measurement units (IMUs) and a new data-processing scheme based on continuous wavelet transforms to address sources of error in the IMU measurements. The headband performance was evaluated in the laboratory on an anthropomorphic test device, which was impacted with a soccer ball to replicate soccer heading. When comparing the measured peak rotational velocities (PRV) and peak rotational accelerations (PRA) between the reference sensors and the headband for impacts to the front of the head, the correlation coefficients (r) were 0.80 and 0.63, and the normalized root mean square error (NRMSE) values were 0.20 and 0.28, respectively. However, when considering all impact locations, r dropped to 0.42 and 0.34 and NRMSE increased to 0.5 and 0.41 for PRV and PRA, respectively. This new instrumented headband improves upon previous headband designs in reconstructing the rotational head kinematics resulting from frontal soccer ball impacts, providing a potential alternative to instrumented mouthguards.
评论: 22 pages, 8 figures, 1 table
主题: 信号处理 (eess.SP)
引用方式: arXiv:2504.01939 [eess.SP]
  (或者 arXiv:2504.01939v1 [eess.SP] 对于此版本)
  https://doi.org/10.48550/arXiv.2504.01939

提交历史

来自:Anu Tripathi
[v1] 2025年4月3日 星期四 00:00:00 UTC
全文链接:

访问论文:

  • 查看原文 PDF
  • 中文 PDF (下批处理)
  • EasyResearch 阅读
  • TeX 源文件
  • arXiv 原文页面
查看许可

当前浏览上下文:

eess.SP
< 上一篇   |   下一篇 >
最新 | 近期 | 2025-04
切换浏览方式为:
eess

参考文献与引用

  • Google Scholar
  • Semantic Scholar

BibTeX 格式的引用

EasyXiv 是独立运营的中文预印本站点,与 arXiv 及康奈尔大学无隶属关系。 论文元数据与原文来自 arXiv.org,感谢 arXiv 提供开放获取服务。
关于 · 帮助 · 投稿指南 · 版权 · arXiv 原站