物理学 > 原子物理
[提交于 2025年7月18日]
标题:Transportable strontium lattice clock with $4 \times 10^{-19}$ blackbody radiation shift uncertainty
查看原文 PDF 中文 PDF (下批处理) EasyResearch 阅读摘要:We describe a transportable optical lattice clock based on the $^1\mathrm{S}_0 \rightarrow {^3\mathrm{P}_0}$ transition of lattice-trapped $^{87}$Sr atoms with a total systematic uncertainty of $2.1 \times 10^{-18}$. The blackbody radiation shift, which is the leading systematic effect in many strontium lattice clocks, is controlled at the level of $4.0 \times 10^{-19}$, as the atoms are interrogated inside a well-characterised, cold thermal shield. Using a transportable clock laser, the clock reaches a frequency instability of about $5 \times 10^{-16}/\sqrt{\tau/\mathrm{s}}$, which enables fast reevaluations of systematic effects. By comparing this clock to the primary caesium fountain clocks CSF1 and CSF2 at Physikalisch-Technische Bundesanstalt, we measure the clock transition frequency with a fractional uncertainty of $1.9\times 10^{-16}$, in agreement with previous results. The clock was successfully transported and operated at different locations. It holds the potential to be used for geodetic measurements with centimetre-level or better height resolution and for accurate inter-institute frequency comparisons.
提交历史
来自:Ingo Nosske[v1] 2025年7月18日 星期五 15:57:32 UTC