标题:Ultra-Sharp Upright Photon Radiotherapy via Low Energy Extended Distance: An Alternative to FLASH for high flux Sources

作者:Lloyd E Kamole Ghomsi, Clinton Gibson, Veng-Jean Heng, Ramish M Ashraf, Lawrie Skinner
查看论文《Ultra-Sharp Upright Photon Radiotherapy via Low Energy Extended Distance: An Alternative to FLASH for high flux Sources》的 PDF
查看中文 PDF 查看原文 PDF EasyResearch 阅读
摘要:Standard 6 megavolt (MV) radiotherapy is limited by source size and secondary electron range to minimum radiological penumbra widths of ~2-3 mm. This study investigates sharper beams via upright radiotherapy with lower energies and extended source-to-patient distances. A 2.5 MV beam from a clinical linac was delivered at a 4 m source-to-phantom distance (2.5 MV-ED). Lateral profiles and percent depth doses were measured in a solid water phantom with radiochromic film and an ion chamber. These single beam measurements were used to benchmark TOPAS Monte Carlo simulations. The validated 2.5 MV-ED model was then used to simulate upright deliveries with a conical beam geometry. These simulations were compared against equivalent plans generated for standard 6 MV-FFF coplanar deliveries at 1 m from the source. The 2.5 MV-ED single 28x28 mm^2 beam produced a measured 80%-20% penumbra of 1.0 +-0.1 mm, compared to 2.4 mm Jaw-defined penumbra for a standard 6MV-FFF beam. The doses at 10 cm depth were 52% vs. 56%, and the surface doses were 22% vs. 38% for the 2.5MV-ED and standard 6 MV-FFF respectively. Conical geometry Monte-Carlo simulations using the 2.5 MV-ED beams demonstrated significantly sharper composite dose fall-off in all cardinal directions compared to coplanar 6MV-FFF plans. For a spatially fractionated lattice example plan with 5 mm diameter high dose spheres, the 2.5 MV-ED conical approach achieved a peak-to-valley dose ratio of 4.5-5.2, compared to the 2.6-2.9 achievable with a standard 6 MV-FFF clinical system. Low-energy, extended-distance photon beams can provide sharper penumbra and lower surface dose, while maintaining comparable depth-dose penetration as standard 6 MV setups. Combined with upright patient setups, ultra-sharp dose distributions with enhanced treatment conformity, reduced toxicity, and higher-fidelity dose modulation are possible.
主题: 医学物理 (physics.med-ph); 加速器物理 (physics.acc-ph); 应用物理 (physics.app-ph)
引用方式: arXiv:2603.07103 [physics.med-ph]
  (或者 arXiv:2603.07103v1 [physics.med-ph] 对于此版本)
  https://doi.org/10.48550/arXiv.2603.07103

提交历史

来自:Lawrie Skinner
[v1] 2026年3月7日 星期六 08:27:09 UTC
全文链接:

访问论文:

  • 查看中文 PDF
  • 查看原文 PDF
  • EasyResearch 阅读
  • TeX 源文件
  • arXiv 原文页面
查看许可

当前浏览上下文:

physics.med-ph
< 上一篇   |   下一篇 >
最新 | 近期 | 2026-03
切换浏览方式为:
physics
physics.acc-ph
physics.app-ph

参考文献与引用

  • Google Scholar
  • Semantic Scholar

BibTeX 格式的引用

EasyXiv 是独立运营的中文预印本站点,与 arXiv 及康奈尔大学无隶属关系。 论文元数据与原文来自 arXiv.org,感谢 arXiv 提供开放获取服务。
关于 · 帮助 · 投稿指南 · 版权 · arXiv 原站