标题:Commissioning and clinical outcome assessment of a novel surface-guided radiation therapy (SGRT) system at a C-Arm linear accelerator
查看原文 PDF 中文 PDF (下批处理) EasyResearch 阅读摘要:Surface-guided radiation therapy (SGRT) is now widely used for radiation-dose-free, marker-free patient positioning in modern radiotherapy. We commissioned and clinically implemented a novel SGRT system, LUNA 3D (LAP, Lueneburg, Germany), featuring browser-based operation, GPU-accelerated surface reconstruction, frame rates above 12 Hz, a large field of view, and virtual laser projection. Commissioning included tests of temperature drift, reproducibility, translational and rotational shift accuracy, gantry-related camera occlusion, agreement with cone-beam CT (CBCT), and end-to-end dosimetric performance. Results were evaluated using both an SGRT-acquired reference surface and a CT-derived external surface. Temperature drift remained below 0.4 mm on all axes. With the SGRT reference, maximum deviations were at most 0.3 mm translationally and 0.2 degrees rotationally; with the CT-derived reference, translational deviations increased to 0.8 mm, consistent with systematic bias from the reference surface. Agreement between LUNA 3D and CBCT was within 1.0 mm, and end-to-end testing showed CBCT residuals of 0.9-1.3 mm with 1.2% dosimetric deviation. All results satisfied ESTRO-ACROP guideline criteria. Clinical evaluation of 192 breast and 259 pelvic treatment datasets showed significant improvements after implementation: the breast 3D translational vector decreased by 28.7% from 7.00 +/- 4.35 mm to 4.99 +/- 2.75 mm, and the pelvic 3D rotational vector decreased by 24.0% from 2.31 +/- 0.96 degrees to 1.76 +/- 0.67 degrees (both p < 0.001). These results establish LUNA 3D as a reliable SGRT system that improves routine patient positioning accuracy.
提交历史
来自:Hui Khee Looe[v1] 2026年4月1日 星期三 07:57:38 UTC (12,436 KB)