标题:Commissioning and clinical outcome assessment of a novel surface-guided radiation therapy (SGRT) system at a C-Arm linear accelerator

作者:Hui Khee Looe, Niklas Felix Hendrik Bartner, Björn Poppe, Kay C. Willborn
查看论文《Commissioning and clinical outcome assessment of a novel surface-guided radiation therapy (SGRT) system at a C-Arm linear accelerator》的 PDF
查看原文 PDF 中文 PDF (下批处理) EasyResearch 阅读
摘要:Surface-guided radiation therapy (SGRT) is now widely used for radiation-dose-free, marker-free patient positioning in modern radiotherapy. We commissioned and clinically implemented a novel SGRT system, LUNA 3D (LAP, Lueneburg, Germany), featuring browser-based operation, GPU-accelerated surface reconstruction, frame rates above 12 Hz, a large field of view, and virtual laser projection. Commissioning included tests of temperature drift, reproducibility, translational and rotational shift accuracy, gantry-related camera occlusion, agreement with cone-beam CT (CBCT), and end-to-end dosimetric performance. Results were evaluated using both an SGRT-acquired reference surface and a CT-derived external surface. Temperature drift remained below 0.4 mm on all axes. With the SGRT reference, maximum deviations were at most 0.3 mm translationally and 0.2 degrees rotationally; with the CT-derived reference, translational deviations increased to 0.8 mm, consistent with systematic bias from the reference surface. Agreement between LUNA 3D and CBCT was within 1.0 mm, and end-to-end testing showed CBCT residuals of 0.9-1.3 mm with 1.2% dosimetric deviation. All results satisfied ESTRO-ACROP guideline criteria. Clinical evaluation of 192 breast and 259 pelvic treatment datasets showed significant improvements after implementation: the breast 3D translational vector decreased by 28.7% from 7.00 +/- 4.35 mm to 4.99 +/- 2.75 mm, and the pelvic 3D rotational vector decreased by 24.0% from 2.31 +/- 0.96 degrees to 1.76 +/- 0.67 degrees (both p < 0.001). These results establish LUNA 3D as a reliable SGRT system that improves routine patient positioning accuracy.
主题: 医学物理 (physics.med-ph)
引用方式: arXiv:2604.00589 [physics.med-ph]
  (或者 arXiv:2604.00589v1 [physics.med-ph] 对于此版本)
  https://doi.org/10.48550/arXiv.2604.00589
相关 DOI: https://doi.org/10.1002/acm2.70707

提交历史

来自:Hui Khee Looe
[v1] 2026年4月1日 星期三 07:57:38 UTC (12,436 KB)
全文链接:

访问论文:

  • 查看原文 PDF
  • 中文 PDF (下批处理)
  • EasyResearch 阅读
  • TeX 源文件
  • arXiv 原文页面
查看许可

当前浏览上下文:

physics.med-ph
< 上一篇   |   下一篇 >
最新 | 近期 | 2026-04
切换浏览方式为:
physics

参考文献与引用

  • Google Scholar
  • Semantic Scholar

BibTeX 格式的引用

EasyXiv 是独立运营的中文预印本站点,与 arXiv 及康奈尔大学无隶属关系。 论文元数据与原文来自 arXiv.org,感谢 arXiv 提供开放获取服务。
关于 · 帮助 · 投稿指南 · 版权 · arXiv 原站