物理学 > 医学物理

arXiv:2607.25440 (physics)
[提交于 2026年7月28日 (v1),最后修订于 2026年11月25日 (此版本, v2)]

标题:Liver Lesion Detection with Task-Specific Organ-Level Tube Current Modulation: A Retrospective Reader Study

作者:Maria Jose Medrano, Sen Wang, Kaylee W. Fang, Jennie Cao, Liyan Sun, Luyao Shen, Andy Shon, Luke Yoon, Grant M. Stevens, Adam S. Wang, Justin R. Tse
查看论文《Liver Lesion Detection with Task-Specific Organ-Level Tube Current Modulation: A Retrospective Reader Study》的 PDF
查看原文 PDF 中文 PDF (下批处理) EasyResearch 阅读
摘要:Rationale and Objectives: To evaluate whether task-specific organ-level TCM improves hepatic lesion detection and abdominal image quality compared with uniform dose reduction at equivalent effective dose. Materials and Methods: In this single-center, IRB-approved, retrospective study, 22 oncology patients with 68 liver lesions (38 malignant, 30 benign) underwent chest-abdomen-pelvis CT. Using projection-domain noise emulation, five series were generated per patient: original acquisition (Orig-100); uniform reduction to 40% (UD-40) and 60% (UD-60) of effective dose; and organ-level TCM at matched 40% (Opt-40) and 60% (Opt-60). Three abdominal radiologists independently detected and classified lesions as benign or malignant and rated image quality on a 5-point Likert scale. Per-lesion sensitivity and image-quality scores were compared using the McNemar and Wilcoxon signed-rank tests, respectively. Malignant lesion-size thresholds (x50, x90) were estimated by logistic regression. Results: At 60% dose, Opt-60 malignant sensitivity was comparable to Orig-100 (82% vs 80%) and higher than UD-60 (67%). At 40% dose, Opt-40 exceeded UD-40 (68% vs 52%). All lesion detection improved similarly (Opt vs UD: 64% vs 54% at 40%; 77% vs 65% at 60%), with Opt-60 comparable to full dose (77% vs 76%). All Opt-versus-UD detection differences were significant (p <= 0.003). Opt-60 lesion-size thresholds (x50 = 3.3 mm; x90 = 15.7 mm) matched Orig-100 (3.7 mm, 16.8 mm). Abdominal image quality was higher with Opt than UD at both doses. Conclusion: At equivalent effective dose, organ-level TCM improved liver lesion detection over uniform reduction, with Opt-60 comparable to full dose. These findings support organ-level optimization for dose reduction while preferentially preserving image quality for liver metastasis detection.
主题: 医学物理 (physics.med-ph)
引用方式: arXiv:2607.25440 [physics.med-ph]
  (或者 arXiv:2607.25440v2 [physics.med-ph] 对于此版本)
  https://doi.org/10.48550/arXiv.2607.25440

提交历史

来自:Maria Jose Medrano Matamoros
[v1] 2026年7月28日 星期二 08:32:51 UTC (5,786 KB)
[v2] 2026年11月25日 星期三 01:22:32 UTC (2,277 KB)
全文链接:

访问论文:

  • 查看原文 PDF
  • 中文 PDF (下批处理)
  • EasyResearch 阅读
  • TeX 源文件
  • arXiv 原文页面
查看许可

当前浏览上下文:

physics.med-ph
< 上一篇   |   下一篇 >
最新 | 近期 | 2026-11
切换浏览方式为:
physics

参考文献与引用

  • Google Scholar
  • Semantic Scholar

BibTeX 格式的引用

EasyXiv 是独立运营的中文预印本站点,与 arXiv 及康奈尔大学无隶属关系。 论文元数据与原文来自 arXiv.org,感谢 arXiv 提供开放获取服务。
关于 · 帮助 · 投稿指南 · 版权 · arXiv 原站