标题:Evaluation of Image Noise from Computed Tomography in an Organ-Specific Anthropomorphic Phantom

作者:Lucas McCullum, Mojtaba Zarei, Andy Tsai, Michael J. Callahan, Peng Sun, Da Zhang
查看论文《Evaluation of Image Noise from Computed Tomography in an Organ-Specific Anthropomorphic Phantom》的 PDF
查看原文 PDF 中文 PDF (下批处理) EasyResearch 阅读
摘要:Computed tomography (CT) is used ubiquitously for diagnostic imaging, with novel iterative reconstructions such as hybrid iterative reconstruction with partial model-based component (ASiR-V) and deep learning image reconstruction (DLIR) increasingly utilized. In conjunction, CT vendors have developed a noise index (NI) to reduce radiation dose during protocol development. However, it is unclear how these reconstruction algorithms reduce noise as NI changes, particularly across multiple organs. The goal of this study was to utilize an anthropomorphic phantom to evaluate image noise against the prescribed NI across different organs, determine if DLIR reduces image noise compared to ASiR-V, and use this relationship to determine the maximum volume CT dose index (CTDIvol) reduction at equivalent noise levels. Repeat images were acquired on a GE HealthCare Revolution Apex CT scanner across NI values in an anthropomorphic adult-sized phantom from Kyoto Kagaku with ASiR-V and DLIR. Organs-of-interest were contoured using TotalSegmentator in 3D Slicer. The correlation between the CTDIvol and NI was determined. The standard deviation of the difference in the repeat images was used as a noise metric (NM), correlated with NI, used to determine the NI at equivalent NM between ASiR-V and DLIR, then applied to determine the dose reduction and difference in NM. The relationship between NI and the resulting CTDIvol was found to be proportional to a power law of exponent -2.23. The maximum potential absolute dose reduction from using DLIR was found in the liver at 1.3 mGy with an NI of 7.9. Across all organs, the relative dose reduction increased, the NM-to-NI ratio decreased, and the noise reduction increased across NI. The relationship between image noise and the prescribed NI differed between ASiR-V and DLIR and varied by organ, with model-based dose reductions of up to 1.3 mGy in the liver.
主题: 医学物理 (physics.med-ph)
引用方式: arXiv:2611.07357 [physics.med-ph]
  (或者 arXiv:2611.07357v1 [physics.med-ph] 对于此版本)
  https://doi.org/10.48550/arXiv.2611.07357

提交历史

来自:Lucas McCullum
[v1] 2026年11月5日 星期四 02:50:41 UTC (1,475 KB)
全文链接:

访问论文:

  • 查看原文 PDF
  • 中文 PDF (下批处理)
  • EasyResearch 阅读
  • TeX 源文件
  • arXiv 原文页面
查看许可

当前浏览上下文:

physics.med-ph
< 上一篇   |   下一篇 >
最新 | 近期 | 2026-11
切换浏览方式为:
physics

参考文献与引用

  • Google Scholar
  • Semantic Scholar

BibTeX 格式的引用

EasyXiv 是独立运营的中文预印本站点,与 arXiv 及康奈尔大学无隶属关系。 论文元数据与原文来自 arXiv.org,感谢 arXiv 提供开放获取服务。
关于 · 帮助 · 投稿指南 · 版权 · arXiv 原站