标题:Design and Performance Evaluation of a Quad-panel Dual-isotope PET Imaging System

作者:Yu Shi, Tengjiao Wang, Sen Yang, Bo Wen, Meng Niu, Yirong Wang, Risheng Yang, Yanyun Liu, Wei Ding, Xiangyu Liu, Fei Kang, Shouping Zhu
查看论文《Design and Performance Evaluation of a Quad-panel Dual-isotope PET Imaging System》的 PDF
查看原文 PDF 中文 PDF (下批处理) EasyResearch 阅读
摘要:In simultaneous dual-isotope PET, prompt-gamma-based separation is challenged by the low statistics of triple coincidences and the limited flexibility of coincidence logic for multi-energy event extraction. We designed and evaluated a compact quad-panel dual-isotope PET system without dedicated prompt gamma detectors. The system comprised four orthogonally arranged detector panels, each providing an effective area of 107.0 * 145.5 mm. GPU-accelerated software coincidence processing extracted double- and prompt-gamma-tagged triple coincidences from raw single-photon data. Triple-coincidence images were reconstructed with correlated-random-triple correction using the consistent delayed expectation-maximization method, followed by image-domain isotope separation using a spatially varying scaling factor. System performance was evaluated using phantom and in vivo studies, with emphasis on 68Ga/18F imaging because of the low prompt gamma yield. The system achieved energy resolutions of 15.7%, 12.9% and 10.2% at 511, 1077, and 1275 keV, and a timing resolution of 1.35 ns. In the 68Ga Derenzo, 1.6-mm rods were resolved in double-coincidence images, with 1.2-mm rods partially distinguishable, while 1.6-mm rods remained discernible in triple-coincidence images. The mean prompt gamma detection efficiencies along the three axes ranged from 8.19% to 9.31%. The system sensitivity was 1.12% for doubles with an energy window of 350-650 keV, and 0.068% for triples after random correction with a dual-energy window. The scatter fraction was 27.81%, and the NECR reached 83 kcps at 9.38 MBq for doubles and 106 cps at 21.42 MBq for triples. Phantom and in vivo studies demonstrated successful separation of 18F/68Ga and 18F/124I tracer distributions. By maintaining high-resolution and quantitative imaging performance, the system provides a preclinical platform for multitracer PET studies.
主题: 医学物理 (physics.med-ph)
引用方式: arXiv:2611.11160 [physics.med-ph]
  (或者 arXiv:2611.11160v1 [physics.med-ph] 对于此版本)
  https://doi.org/10.48550/arXiv.2611.11160

提交历史

来自:Yu Shi
[v1] 2026年11月8日 星期日 05:59:33 UTC (6,905 KB)
全文链接:

访问论文:

  • 查看原文 PDF
  • 中文 PDF (下批处理)
  • EasyResearch 阅读
  • TeX 源文件
  • arXiv 原文页面
查看许可

当前浏览上下文:

physics.med-ph
< 上一篇   |   下一篇 >
最新 | 近期 | 2026-11
切换浏览方式为:
physics

参考文献与引用

  • Google Scholar
  • Semantic Scholar

BibTeX 格式的引用

EasyXiv 是独立运营的中文预印本站点,与 arXiv 及康奈尔大学无隶属关系。 论文元数据与原文来自 arXiv.org,感谢 arXiv 提供开放获取服务。
关于 · 帮助 · 投稿指南 · 版权 · arXiv 原站