物理学 > 仪器与探测器

arXiv:2611.12369 (physics)
[提交于 2026年11月9日]

标题:VNA-Only Calibration of Waveguide Thermistor Mounts with Self-Consistent Determination of the Thermistor-Port Impedance

作者:Patricio Greco
查看论文《VNA-Only Calibration of Waveguide Thermistor Mounts with Self-Consistent Determination of the Thermistor-Port Impedance》的 PDF
查看原文 PDF 中文 PDF (下批处理) EasyResearch 阅读
摘要:A method is presented for determining the calibration factor and the effective efficiency of waveguide thermistor mounts from vector-network-analyzer reflection data alone, without a microcalorimeter, extending the three reflection-coefficient method of the coaxial case. In waveguide the thermistor is coupled to the line through a matching structure and the reference impedance of the thermistor port, $Z_0$, is not known a priori; it is inferred through a power-consistency relation between pairs of states, under the convention that it is real. The method reproduces the factory certificates (U = 2 %) of four K486A mounts within 1.0 % rms and the PTB-traceable values (U = 0.85 to 0.98 %) of a fifth within 0.34 % rms; its expanded measurement uncertainty is 0.5 to 1.16 % (k = 2) for the best specimen and below 2 % for the five. A mount with its waveguide-to-coaxial adapter is calibrated directly, without characterizing the adapter, from the model factor of the mount and the geometric factor measured through the adapter. The scattering matrix implied by the solution is not passive; a minimum-norm correction of the thermistor impedances along the two directions invisible to the calibration factor - the ratios of the efficiencies of the three states - brings it to the floor of passivity, where the efficiency depends on the load only through its mismatch factor, leaves the result unchanged, and gives the sensitivity of the efficiency to the bias resistance in closed form: below $4\times10^{-5}$ $\Omega^{-1}$ in band mean, proportional to the mismatch of the operating state and to the loss of the mount.
评论: 60 pages, 21 figures. Submitted to Metrologia (IOP Publishing / BIPM), November 2026
主题: 仪器与探测器 (physics.ins-det); 信号处理 (eess.SP)
引用方式: arXiv:2611.12369 [physics.ins-det]
  (或者 arXiv:2611.12369v1 [physics.ins-det] 对于此版本)
  https://doi.org/10.48550/arXiv.2611.12369
相关 DOI: https://doi.org/10.5281/zenodo.24038232

提交历史

来自:Patricio Greco
[v1] 2026年11月9日 星期一 13:15:48 UTC (1,427 KB)
全文链接:

访问论文:

  • 查看原文 PDF
  • 中文 PDF (下批处理)
  • EasyResearch 阅读
  • TeX 源文件
  • arXiv 原文页面
查看许可

当前浏览上下文:

physics.ins-det
< 上一篇   |   下一篇 >
最新 | 近期 | 2026-11
切换浏览方式为:
physics
eess.SP

参考文献与引用

  • Google Scholar
  • Semantic Scholar

BibTeX 格式的引用

EasyXiv 是独立运营的中文预印本站点,与 arXiv 及康奈尔大学无隶属关系。 论文元数据与原文来自 arXiv.org,感谢 arXiv 提供开放获取服务。
关于 · 帮助 · 投稿指南 · 版权 · arXiv 原站