arXiv:2611.33924 (physics)
[提交于 2026年11月25日]

标题:Monte Carlo-based secondary dose check platform for the world's first ultra-compact upright proton therapy system

作者:Toghrul Alakbarzade, Emir Musa Olgun, Yuan James Rao, Billy W Loo Jr, Adam Johansson, Yong Yang, Murat Surucu, Mehmet Burcin Unlu, Serdar Charyyev
查看论文《Monte Carlo-based secondary dose check platform for the world's first ultra-compact upright proton therapy system》的 PDF
查看原文 PDF 中文 PDF (下批处理) EasyResearch 阅读
摘要:Objective. This work aimed to develop and validate an independent Monte Carlo (MC) secondary dose check of the RayStation treatment planning system (TPS) for the Mevion S250-FIT, the world's first ultra-compact upright proton therapy system. Approach. The treatment nozzle, including the energy modulation system of boron carbide and polycarbonate plates and the adaptive aperture, was modeled in the TOPAS (TOol for PArticle Simulation) MC code, and the beam model was commissioned against integrated depth dose curves, in-air spot profiles, and dose per monitor unit measured at 12 energy settings. Based on this model, an in-house converter, RayStation2TOPAS, recalculates the TPS plans by reconstructing each pencil beam spot, the aperture for each energy layer, and the seated patient's rotation. In total, 15 water phantom plans and 25 clinical plans were recalculated and compared with the TPS by the point dose, with a tolerance of 5%, and by three-dimensional gamma analysis at 3%/2mm, with a pass rate of at least 90%. Main results. The simulated proton range agreed with the measurement within 0.25mm, the spot full width at half maximum within 6.8%, and the dose per monitor unit within 1.3%. The point dose differed from the TPS by -0.76% to +1.98% in the water phantom plans and by -3.03% to +3.99% in the clinical plans, with gamma pass rates of 96.39% to 99.93% and 90.08% to 99.98%, so every plan passed both tests. Significance. To our knowledge, this is the first independent MC secondary dose check for the S250-FIT, and it complements measurement-based patient-specific quality assurance of its plans.
主题: 医学物理 (physics.med-ph)
引用方式: arXiv:2611.33924 [physics.med-ph]
  (或者 arXiv:2611.33924v1 [physics.med-ph] 对于此版本)
  https://doi.org/10.48550/arXiv.2611.33924

提交历史

来自:Serdar Charyyev
[v1] 2026年11月25日 星期三 07:44:38 UTC (2,163 KB)
全文链接:

访问论文:

  • 查看原文 PDF
  • 中文 PDF (下批处理)
  • EasyResearch 阅读
  • TeX 源文件
  • arXiv 原文页面
查看许可

当前浏览上下文:

physics.med-ph
< 上一篇   |   下一篇 >
最新 | 近期 | 2026-11
切换浏览方式为:
physics

参考文献与引用

  • Google Scholar
  • Semantic Scholar

BibTeX 格式的引用

EasyXiv 是独立运营的中文预印本站点,与 arXiv 及康奈尔大学无隶属关系。 论文元数据与原文来自 arXiv.org,感谢 arXiv 提供开放获取服务。
关于 · 帮助 · 投稿指南 · 版权 · arXiv 原站