物理学 > 仪器与探测器
[提交于 2002年10月14日]
标题:Low noise temperature control : application to an active cavity radiometer
查看原文 PDF 中文 PDF (下批处理) EasyResearch 阅读摘要:We have designed low noise temperature sensing and control units with the objective of using them for the fabrication of far infrared active cavity radiometers. The sensing unit, first characterized at 300K using industrial platinium resistance thermometers, has a noise level of 25E-6 to 30E-6 Keff for a 3 hours measuring time and in a 1Hz bandwidth. Using YBCO superconducting thermometers, the noise level goes down to 2.5E-6 Keff, which is strongly limited by excess 1/f noise in the YBCO film at the superconducting transition. The sample holder used in the 90K experiments is built with an auxiliary heating resistor, which enables an easy and accurate identification of the electrothermal model, even in the closed loop operation. Based on a design previously published by NIST, we estimate from these experimental results that the overall noise limitations of radiometers could be lowered by one order of magnitude.
提交历史
来自:B. Guillet[v1] 2002年10月14日 星期一 11:43:13 UTC