标题:Low noise temperature control : application to an active cavity radiometer

作者:B. Guillet, D. Robbes, L. Mechin
查看论文《Low noise temperature control : application to an active cavity radiometer》的 PDF
查看原文 PDF 中文 PDF (下批处理) EasyResearch 阅读
摘要:We have designed low noise temperature sensing and control units with the objective of using them for the fabrication of far infrared active cavity radiometers. The sensing unit, first characterized at 300K using industrial platinium resistance thermometers, has a noise level of 25E-6 to 30E-6 Keff for a 3 hours measuring time and in a 1Hz bandwidth. Using YBCO superconducting thermometers, the noise level goes down to 2.5E-6 Keff, which is strongly limited by excess 1/f noise in the YBCO film at the superconducting transition. The sample holder used in the 90K experiments is built with an auxiliary heating resistor, which enables an easy and accurate identification of the electrothermal model, even in the closed loop operation. Based on a design previously published by NIST, we estimate from these experimental results that the overall noise limitations of radiometers could be lowered by one order of magnitude.
评论: 19 pages, 12 figures, Accepted to Rev. Sci. Instrum
主题: 仪器与探测器 (physics.ins-det)
引用方式: arXiv:physics/0210059 [physics.ins-det]
  (或者 arXiv:physics/0210059v1 [physics.ins-det] 对于此版本)
  https://doi.org/10.48550/arXiv.physics/0210059
相关 DOI: https://doi.org/10.1063/1.1525872

提交历史

来自:B. Guillet
[v1] 2002年10月14日 星期一 11:43:13 UTC
全文链接:

访问论文:

  • 查看原文 PDF
  • 中文 PDF (下批处理)
  • EasyResearch 阅读
  • TeX 源文件
  • arXiv 原文页面
查看许可

当前浏览上下文:

physics.ins-det
< 上一篇   |   下一篇 >
最新 | 近期 | 2002-10
切换浏览方式为:
physics

参考文献与引用

  • Google Scholar
  • Semantic Scholar

BibTeX 格式的引用

EasyXiv 是独立运营的中文预印本站点,与 arXiv 及康奈尔大学无隶属关系。 论文元数据与原文来自 arXiv.org,感谢 arXiv 提供开放获取服务。
关于 · 帮助 · 投稿指南 · 版权 · arXiv 原站